IP Annals of Prosthodontics and Restorative Dentistry

Print ISSN: 2581-4796

Online ISSN: 2581-480X

IP Annals of Prosthodontics and Restorative Dentistry (APRD) open access, peer-reviewed quarterly journal publishing since 2015 and is published under the Khyati Education and Research Foundation (KERF), is registered as a non-profit society (under the society registration act, 1860), Government of India with the vision of various accredited vocational courses in healthcare, education, paramedical, yoga, publication, teaching and research activity, with the aim of faster and better dissemination of knowledge, we will be publishing the article more...

Article type

Original Article


Article page

97-103


Authors Details

Khalid Abograra Ahmad Shaban, Fatma Mahanna, Nesreen El Mekawy*


Article Metrics


View Article As

 


Downlaod Files

   






Article statistics

Viewed: 1569

PDF Downloaded: 901


Fitness evaluation of mandibular kennedy class I BioHPP Poly-Ether-Ether ketone removable partial denture fabricated by various techniques


Original Article

Author Details : Khalid Abograra Ahmad Shaban, Fatma Mahanna, Nesreen El Mekawy*

Volume : 5, Issue : 4, Year : 2019

Article Page : 97-103

https://doi.org/10.18231/j.aprd.2019.023



Suggest article by email

Get Permission

Abstract

Purpose: The purpose of this study was to evaluate the fitness of mandibular Kennedy class I PEEK removable partial dentures fabricated with various techniques.
Materials and Methods: Six patients were selected for this study with maxillary completely edentulous arch against mandibular distal extension ridges posterior to canine teeth. each patient was randomly given the following frameworks in a crossover design; Group I: All patients were received PEEK frameworks fabricated by milling via CAD-CAM technique (PEEK disc were used). Group II: All patients were received PEEK frameworks fabricated by injection molding technique (Granular form of medical grade PEEK). For each framework (milled & injected), the fitness evaluation were done by measuring the gap between the framework and oral structures by evaluation the thickness of light body poly vinyl- siloxane impression material.
Results: The mean overall and site specific values of Gap distance beneath frameworks fabricated by CAD-CAM technique was significantly higher than the mean values of Gap distance in frameworks fabricated by injection technique.
Conclusion: Within the limitations of this short-term clinical study it was concluded that: 1) The fitness of both techniques of the RPD PEEK BioHPP frameworks manufacture were clinically accepted by the participants. 2) The RPD PEEK BioHPP frameworks manufactured by CAD / CAM was less fitness values than BioHPP frameworks constructed by Injection molding technique.

Keywords: Kennedy class I, Polyetheretherketone, CAD/CAM technique, Injection mold technique, Removable partial denture fitness..


How to cite : Shaban K A A, Mahanna F , Mekawy N E, Fitness evaluation of mandibular kennedy class I BioHPP Poly-Ether-Ether ketone removable partial denture fabricated by various techniques. IP Ann Prosthodont Restor Dent 2019;5(4):97-103

This is an Open Access (OA) journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.